Transmisor de punto de rocío DMT143 DRYCAP® de Vaisala

    Especificaciones

     Technical Specifications

    Dew point temperature measurement

     

    Measurement range (typical)
    Accuracy in air or N2
    Sensor

    -60 ... +60 °C (-76 ... +140 °F) Td
    ±2 °C (±3.6 °F) Td
    Vaisala DRYCAP® 180D

    ppm volume concentration

     

    Measurement range
    Accuracy at +20 °C (+68 °F),
    1 bar

    m
        1 ppm + 20 % of reading

    Operating environment

     

    Measured gases
    Temperature
    Relative humidity
    Pressure

    s
    -40 ...+60 °C (-40 ... +140 °F)
    0 ... 100 %RH
    0 ... 50 bara(725 psia)

    Outputs

    ​Analog output (scalable) ​4 ... 20 mA (3-wire)
    0 ... 1 V/5 V
    1 ... 5 V

    Digital output

    RS-485, non-isolated
    Vaisala Industrial Protocol

    Mechanics

     

    Weight

    Material (housing)
    Housing classification

    z)
    NPT-thread version 100g (3.5oz)
    stainless steel body (AISI316L)
    IP66 (NEM

    A 4)

     

    .

    Related Services